Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the stent delivery wire (sdw) was found kinked/bent and part of the stent was seen to remaining in the sheath.The stent was also found broken/fractured.Functional testing could not be performed due to the damage noted to the device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The damage noted to the device confirms the reported issue and it can be presumed that the damages occurred during the procedure and therefore, the as reported as well as the analyzed issues were assigned procedural factors as this complaint appears to be associated with a product that met stryker and design and manufacturing specifications and was used in accordance with the dfu.
|