Model Number S606 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Use of Device Problem (1670)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Code Available (3191); Insufficient Information (4580)
|
Event Date 07/24/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
No additional information is available.If additional information becomes available the report will be updated at that time.
|
|
Event Description
|
It was reported that the pacemaker was showing a message it was in magnet mode when there was no magnet present.Boston scientific technical services (ts) was consulted and walked the field representative through attempting to reset the magnet operating from asynchronous to off but it would not allow any programming.Electrogram (egms) were also not available.Ts noted that the pacemaker appears to be not performing as intended and recommended replacement.At this time the pacemaker remains in service and no adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Patient code (b)(6) captures the reportable event of surgery.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
|
|
Event Description
|
Additional information as received that prior to the explant procedure, the pacemaker was interrogated and showed normal device function with no magnet mode noted.The battery status showed two years remaining.The field representative consulted boston scientific technical services (ts) who noted that since it was confirmed at the last check that the device was stuck in a magnet mode and the issue could not be resolved with troubleshooting, they could not guarantee that this behavior would not happen again in this device.Ts recommended device replacement.Subsequently the pacemaker was explanted and replaced.No additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that the pacemaker was showing a message it was in magnet mode when there was no magnet present.Boston scientific technical services (ts) was consulted and walked the field representative through attempting to reset the magnet operating from asynchronous to off but it would not allow any programming.Electrogram (egms) were also not available.Ts noted that the pacemaker appears to be not performing as intended and recommended replacement.At this time the pacemaker remains in service and no adverse patient effects were reported.Additional information as received that prior to the explant procedure, the pacemaker was interrogated and showed normal device function with no magnet mode noted.The battery status showed two years remaining.The field representative consulted boston scientific technical services (ts) who noted that since it was confirmed at the last check that the device was stuck in a magnet mode and the issue could not be resolved with troubleshooting, they could not guarantee that this behavior would not happen again in this device.Ts recommended device replacement.Subsequently the pacemaker was explanted and replaced.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Patient code 3191 captures the reportable event of surgery.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Magnet function was tested 15 times.With a magnet applied the pacing rate went to 100 paces per minute with the magnet removed the pacing rate returned to the lower rate limit- normal magnet switch operation was noted during analysis.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
|
Search Alerts/Recalls
|