Model Number PE074F5
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/27/2020
Event Type  malfunction  
Manufacturer Narrative
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.As received, the balloon latex, distal and proximal windings were completely missing and were not returned.No visible damage to the returned syringe and catheter body was found.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of balloon inflation issue could not be confirmed during the product evaluation as the balloon latex, and both the distal and proximal windings were completely missing and were not returned.Upon follow-up with the customer there was no change to the initial report of balloon inflation issue.The customer stated that they did not know that the balloon and windings were missing.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.It is unknown if user or procedural factors may have contributed to the event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon of the swan ganz catheter did not inflate at the balloon inflation test before use.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500249
MDR Report Key10620649
MDR Text Key209633510
Report Number2015691-2020-13804
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/06/2022
Device Model NumberPE074F5
Device Lot Number62975175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/18/2020
Initial Date FDA Received10/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-