The patient was undergoing a coil embolization procedure in the internal iliac artery (iia) using a pod coil and a lantern delivery microcatheter (lantern).During the procedure, the physician encountered resistance while attempting to insert a pod coil into the lantern and the pusher assembly of the pod coil broke; therefore, the pod coil was removed.The procedure was completed using a new pod coil and the same lantern.There was no report of an adverse effect to the patient.
|
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 75.5 cm from its proximal end.The pusher assembly mid-joint was proximal to the introducer sheath friction lock.The embolization coil was intact with its pusher assembly.The introducer sheath was ovalized approximately 3.0 cm from the distal tip.Conclusions: evaluation of the returned pod coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock, and the distal tip of the introducer sheath was ovalized.This damage was likely incidental to the reported complaint and may have occurred during packaging for return to penumbra.Due to the returned damage, the functional testing could not be performed; therefore, the root cause of the reported complaint could not be determined.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|