SMITH & NEPHEW, INC. JOURNEY DCF AP FEM CUT BLK 4; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 74012414 |
Device Problems
Break (1069); Material Separation (1562)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/08/2020 |
Event Type
Injury
|
Event Description
|
It was reported that, during surgery, while instrument was inside the patient, the size 4 5/1 cut guide vibrated and became disassembled during the procedure.It is unknown how the procedure was completed.Surgery was not delayed.The patient was not harmed beyond the described event.
|
|
Manufacturer Narrative
|
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this case reports the femoral cutting block became disassembled during use.Per email communication, all pieces were recovered.It is unknown whether a backup device was needed.There was no patient injury or delay reported.Therefore, since no patient harm is alleged, no further clinical assessment is warranted.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|