The reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A visual inspection observed a damaged bezel and the touchscreen is loose.Signs of impact damage around touch screen.A functional evaluation revealed the unit does boot up and the led checks pass on ports and saline pump, but the touchscreen is blank.The unit was opened and found the display cable is disconnected from the main board.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.There are no indications to suggest that the device/product did not meet specifications upon release into distribution.
|