ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1456Q/86
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problems Syncope (1610); Twiddlers Syndrome (2114)
Event Date 09/14/2020
Event Type  Injury  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 2017865-2020-14775.It was reported that the patient presented to the emergency room with syncope.Upon device interrogation episodes of non-sustained right ventricular over-sensing were apparent.The following day intermittent loss of capture and noise was exhibited by the right ventricular lead when the patient was in the supine position.Loss of left ventricular lead capture was also noted.The patient was scheduled for lead revision.During the procedure it was discovered that the both leads were tangled together due to twiddler's syndrome.The leads were explanted and replaced.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key10627650
MDR Text Key209810160
Report Number2017865-2020-14778
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000080304
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/15/2020
Initial Date FDA Received10/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/22/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR
Patient Outcome(s) Required Intervention;
Patient Age19 YR
-
-