Model Number CMS15-10C-US
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/15/2020
Event Type  malfunction  
Event Description
It was reported that the distal filter failed to capture.A sentinel cerebral protection system was selected for a transcatheter aortic valve replacement procedure.During the procedure, the distal slider#3 broke off the back of the device handle when trying to receive the distal filter.The distal filter was not able to be captured.The device was removed with the distal filter open with no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTINEL CEREBRAL PROTECTION SYSTEM (US)
Type of Device
EMBOLIC PROTECTION DEVICE
Manufacturer (Section D)
CLARET MEDICAL, INC.
1745 copperhill parkway
santa rosa CA 95403
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
160 knowles drive
los gatos CA 95032
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10638077
MDR Text Key210078132
Report Number2134265-2020-13493
Device Sequence Number1
Product Code PUM
UDI-Device Identifier00863229000004
UDI-Public00863229000004
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/12/2022
Device Model NumberCMS15-10C-US
Device Catalogue NumberCMS15-10C-US
Device Lot Number0025851845
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/23/2020
Initial Date FDA Received10/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/12/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-