The product has been received for analysis.This report will be updated upon completion of analysis.The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.During analysis an is4 lead was inserted into the lv lead port.The lead was successfully inserted and set screw was able to be secured.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis was unable to confirm the field allegations during testing.
|