MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIG HP FIXED REF AP BLK SZ1.5; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIG HP FIXED REF AP BLK SZ1.5; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES Back to Search Results
Model Number 9505-02-152
Device Problem Material Deformation (2976)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 09/22/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the surgeon was unable to place a drill pin through the +2 posterior referencing drill hole as the metal surrounding the drill hole has become warped due to repeated use.The metal was blocking the drill pin from going through the drill hole.No surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirms the reported event of deformation.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG HP FIXED REF AP BLK SZ1.5
Type of Device
KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10639121
MDR Text Key210180787
Report Number1818910-2020-21674
Device Sequence Number1
Product Code HTZ
UDI-Device Identifier10603295228516
UDI-Public10603295228516
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9505-02-152
Device Catalogue Number950502152
Device Lot NumberJ0508J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Date Manufacturer Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No