• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 100 SGL KIT - FRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 100 SGL KIT - FRA Back to Search Results
Model Number 27072
Device Problem Pressure Problem (3012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/04/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned to resmed for an engineering investigation.The investigation methods, results and conclusion are not finalized at this stage.When more information is available, a supplemental report will be submitted.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf108) related to pressure measurement.There was no patient harm or serious injury reported as a result of this incident.
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf108) related to pressure measurement.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Performance testing could not reproduce the reported complaint.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to operator error.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 SGL KIT - FRA
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU  2153
MDR Report Key10639918
MDR Text Key210377527
Report Number3007573469-2020-00951
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Type of Report Initial,Followup
Report Date 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27072
Device Catalogue Number27072
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2020
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/10/2021
Distributor Facility Aware Date01/20/2021
Device Age55 MO
Date Report to Manufacturer02/10/2021
Date Manufacturer Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-