Model Number 27072 |
Device Problem
Pressure Problem (3012)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to resmed for an engineering investigation.The investigation methods, results and conclusion are not finalized at this stage.When more information is available, a supplemental report will be submitted.Resmed reference#: (b)(4).
|
|
Event Description
|
It was reported to resmed that an astral device displayed an error message (sf108) related to pressure measurement.There was no patient harm or serious injury reported as a result of this incident.
|
|
Event Description
|
It was reported to resmed that an astral device displayed an error message (sf108) related to pressure measurement.There was no patient harm or serious injury reported as a result of this incident.
|
|
Manufacturer Narrative
|
The astral device was returned to resmed for an investigation.Performance testing could not reproduce the reported complaint.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to operator error.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
|
|