MAUDE Adverse Event Report: SMITH & NEPHEW, INC. HI-LUBRIMET METAL INSERT STD. 6-11/32; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. HI-LUBRIMET METAL INSERT STD. 6-11/32; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL Back to Search Results
Catalog Number 75001072
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Injury (2348); Metal Related Pathology (4530)
Event Date 07/03/2020
Event Type  Injury  
Event Description
It was reported that after 14 years of primary procedure, revision with acetabular reconstruction using a ¿ganz-hakendachs shell¿ and trabecular metal augments combined with an uncemented, modular conical revision stem.Three s+n devices were explanted.
 
Manufacturer Narrative
This complaint investigation is already covered under fda report number: 1020279-2020-05208 and under smith and nephew reference: (b)(4).
 
Manufacturer Narrative
This complaint investigation is already covered under fda report number:(b)(4) and under smith and nephew reference: (b)(4).
 
Manufacturer Narrative
It was reported that 14 years after the implantation, a revision surgery was performed on the left hip due to signs of metalosis.Pre-revision cobalt and chromium ion levels were 214nmol/l each.The removed devices, a hi cup, metal insert, metal ball head and an sl stem, used in treatment, were returned for investigation.The liner was disassembled from the cup and the ball head was removed from the stem after the reception of the patient's consent.Furthermore, it was reported that the right hip was revised as well, captured under (b)(4).The liner shows some minor scratches on the contact area with the ball head.The ball head shows a similar pattern.The shell shows some minor signs of use.The stem has some major scratches on the surface, which most likely originate from the explantation procedure.The cone of the stem shows discoloration on the surface.A technical evaluation was conducted.The linear wear of the ball head and the liner were assessed using a roundness machine for the head and a surface profiler for the cup.The linear wear on the ball head was measured to be 54.2 ¿m, the linear wear on the liner was 53.0 ¿m (combined 107.2 ¿m, 7.6 ¿m/year).The volumetric wear could only be assessed for the ball head and resulted in 20.6 mm3.The presence of a marked wear patch on the acetabular liner shows that edge loading has occurred.Taper analysis for the linear wear of the ball head measured the maximum depth of material loss measured at 56.1 ¿m.Compared to former cases, the findings regarding the wear on the bearing surface is higher than expected.A production documentation of all four implants, including the material certificates, did not show any deviation.The certificates demonstrate that the raw material is a wrought cobaltchromium-molybdenum alloy as specified in the product files.There are no indications that the parts failed to match specification at the time of manufacturing.A complaint history was conducted and it was shown that the reported failure mode is covered by the risk management file in severity and occurrence.The combination of the devices was approved by smith and nephew.The ball head and the liner are no longer available on the market.Abrasion of implant surfaces and development of osteolysis as a reaction to foreign bodies is a possible side effect from hip arthroplasty according to the ifu versions, lit no.12.23, ed.09/05 and lit no.12.07, ed.01/03.Based on the received information a medical assessment was conducted.It is stated that the inclination of the cup is pretty flat with about 24 degrees.According to the surgical technique (lit.No.01773-en, v6, 07/20), the acetabular component is to be impacted with 10-20 degrees anteversion and 40-50 degrees inclination.The provided medical information confirms the reported blood cobalt ion level of 214 nmol/l and chromium ion level of 214nmol/l.The reported loosening of the acetabular component can be seen as a result of metallosis but loosening and the migration of the cup from the reported fall could also accelerate wear, lead to metal debris, and result in metalosis.However, the confirmed edge loading cannot be ruled out as a contributing factor.Therefore, the relationship between the reported event and the devices can be confirmed.The devices in questions, ball head stem and liner, are no longer available on the market.A wear patch was observed on the metal-on-metal joint.The investigation performed demonstrate that the material of all implants is within specification and that the material characteristics did not cause the observed metal wear.The alignment of cup and ball head may have contributed to the formation of a wear patch on the metal-on-metal joint.The root cause is therefore attributed to a known inherent risk.The need for further actions is not indicated.Smith and nephew will monitor these devices for further similar issues.The returned devices will be retained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-LUBRIMET METAL INSERT STD. 6-11/32
Type of Device
PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10645428
MDR Text Key210354884
Report Number1020279-2020-05209
Device Sequence Number1
Product Code JDH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number75001072
Device Lot Number047095/19
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2020
Initial Date Manufacturer Received 09/10/2020
Initial Date FDA Received10/07/2020
Supplement Dates Manufacturer Received03/30/2021
03/30/2021
08/19/2021
Supplement Dates FDA Received04/20/2021
04/20/2021
08/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age65 YR
Patient Weight72
-
-