• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Failure to Charge (1085); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 07oct2020.
 
Event Description
It was reported that the unit alarmed to check the battery.There was no patient involvement.The field service engineer (fse) provided remote support and through troubleshooting determined the backup battery was not charging.The fse instructed the customer to run on battery power per the pvt battery test and if after 20 minutes of running on battery the battery does not charge then to replace the power management board.
 
Manufacturer Narrative
Following the evaluation of the device, the field service engineer (fse) recommended replacing the battery.Multiple attempts were made to follow up with the customer to obtain repair information; however, there was no response.If additional information is received at a later date, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
carlsbad CA
Manufacturer Contact
bill cole
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key10645937
MDR Text Key211474938
Report Number2031642-2020-03628
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-