Catalog Number V041163000000 |
Device Problems
Mechanical Problem (1384); Calibration Problem (2890)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it was reported speed incorrect/too fast/too slow for the vdw.Silver device.The event outcome is unknown as of this mdr evaluation.
|
|
Manufacturer Narrative
|
Battery (voltage collapses under load) defective, replaced.Angle piece (2009) 16578 (plain bearing worn) defective, replaced with 12019 (2008).Micro motor sm13494 and foot control 148467 without errors.Function test and test measurements without errors.Additional information was received indicating that no injury resulted.
|
|