Catalog Number 394945
Device Problems Break (1069); Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 09/16/2020
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.(b)(4).
 
Event Description
It was reported that the bd connecta¿ stopcock was found damaged before use in the packaging.The following information was provided by the initial reporter: "appears to be damaged in the packaging".
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2020-10-13.H6: investigation summary: a device history record review was performed for provided lot number 9127844 and the review did not reveal any detected quality issues during the production process that could have contributed to this reported incident.To aid in the investigation of this issue, the physical sample was returned for evaluation by our quality engineer team.Through examination of the sample, a foreign piece of plastic was observed within the sealed product packaging.It is most likely that this foreign material was used to cover a hose within the manufacturing machinery and the material became loose during the production process.In response to this incident, a quality alert has been issued to all applicable manufacturing personnel to raise awareness for this potential defect on the production floor.H3 other text : see h.10.
 
Event Description
It was reported that the bd connecta¿ stopcock was found damaged before use in the packaging.The following information was provided by the initial reporter: "appears to be damaged in the packaging.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CONNECTA STOPCOCK
Type of Device
STOPCOCK
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
MDR Report Key10646388
MDR Text Key226801466
Report Number9610847-2020-00316
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number394945
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Initial Date Manufacturer Received 09/16/2020
Initial Date FDA Received10/07/2020
Supplement Dates Manufacturer Received11/23/2020
Supplement Dates FDA Received11/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-