• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD Back to Search Results
Model Number MN10450-50A
Device Problem High impedance (1291)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/16/2020
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Related manufacturer reference# 1627487-2020-32268, 1627487-2020-32269, 1627487-2020-32267 it was reported the patient was receiving inadequate stimulation.An impedance check revealed high impedance on two leads.Reprogramming was able to provide sufficient therapy.However, the patient may undergo surgical intervention on a later date.All of the patient's leads are being reported because it is unknown which leads are liable.
 
Event Description
Follow-up information indicated the physician explanted and replaced two leads.This issue is now resolved.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Type of Device
DRG SLIM TIP LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key10646830
MDR Text Key210364768
Report Number1627487-2020-32270
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067027153
UDI-Public05415067027153
Combination Product (y/n)N
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/11/2021
Device Model NumberMN10450-50A
Device Catalogue NumberMN10450-50A
Device Lot Number7034968
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/16/2020
Initial Date FDA Received10/08/2020
Supplement Dates Manufacturer Received10/22/2020
12/01/2020
Supplement Dates FDA Received10/27/2020
12/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DRG LEAD (3); DRG LEAD (3)
Patient Outcome(s) Other;
Patient Weight75
-
-