• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS ANTI-SARS-COV-2; CORONAVIRUS SEROLOGICAL REAGENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS ANTI-SARS-COV-2; CORONAVIRUS SEROLOGICAL REAGENT Back to Search Results
Catalog Number 09203095190
Device Problem False Negative Result (1225)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/14/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The initial reporter stated they received discrepant negative results for one patient sample tested with the elecsys anti-sars-cov-2 assay on a cobas e 411 immunoassay analyzer.No incorrect results were reported outside of the laboratory.It is not known which results were considered to be correct.The sample was initially tested with the elecsys anti-sars-cov-2 assay, resulting in a value of 0.219 coi (non-reactive).The initial testing of the sample was performed <30 days after pcr testing.The sample was repeated with the elecsys anti-sars-cov-2 assay on (b)(6) 2020, resulting in a value of 56.3 coi (reactive).The e411 analyzer serial number is (b)(4).
 
Manufacturer Narrative
The last calibration performed on 12-sep-2020 was ok.Quality controls recovered within range.Upon review of the alarm trace, no relevant alarms were observed.The investigation could not identify a product problem.The cause of the event could not be determined.Product labeling states, "the sensitivity of the elecsys anti-sars-cov-2 assay early after infection is unknown.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-SARS-COV-2
Type of Device
CORONAVIRUS SEROLOGICAL REAGENT
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10648656
MDR Text Key242650999
Report Number1823260-2020-02490
Device Sequence Number1
Product Code QKO
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09203095190
Device Lot Number50726000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/18/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age66 YR
-
-