Catalog Number 09203095190 |
Device Problem
False Negative Result (1225)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in (b)(6).
|
|
Event Description
|
The initial reporter stated they received discrepant negative results for one patient sample tested with the elecsys anti-sars-cov-2 assay on a cobas e 411 immunoassay analyzer.No incorrect results were reported outside of the laboratory.It is not known which results were considered to be correct.The sample was initially tested with the elecsys anti-sars-cov-2 assay, resulting in a value of 0.219 coi (non-reactive).The initial testing of the sample was performed <30 days after pcr testing.The sample was repeated with the elecsys anti-sars-cov-2 assay on (b)(6) 2020, resulting in a value of 56.3 coi (reactive).The e411 analyzer serial number is (b)(4).
|
|
Manufacturer Narrative
|
The last calibration performed on 12-sep-2020 was ok.Quality controls recovered within range.Upon review of the alarm trace, no relevant alarms were observed.The investigation could not identify a product problem.The cause of the event could not be determined.Product labeling states, "the sensitivity of the elecsys anti-sars-cov-2 assay early after infection is unknown.".
|
|