• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problems Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
Patient Problems Sweating (2444); Palpitations (2467); Shaking/Tremors (2515)
Event Date 09/27/2020
Event Type  Injury  
Manufacturer Narrative
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kit were reviewed and the dhrs showed the libre sensor and libre sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A caller reported a customer received an error message, "scan again in 10 minutes", for 2 hours or more with the adc freestyle libre 2 sensor.On (b)(6) 2020, the customer experienced sweating, trembling, and palpitations and was unable to treat themselves.The customer was provided unspecified treatment by his wife for his symptoms.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.The sensor plug was properly seated and no issues were observed with the returned sensor patch.Extracted data from the returned sensor using approved software.Sensor found to be in sensor state 5 (indicating normal termination).Visually inspected the plug assembly and no issues were observed.A sim vivo test was performed and the results were with in specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.Additional information- section g1: (contact office first name, contact office last name, contact office phone number and contact office email) have been updated to (b)(6).
 
Event Description
A caller reported a customer received an error message, "scan again in 10 minutes", for 2 hours or more with the adc freestyle libre 2 sensor.On (b)(6)2020, the customer experienced sweating, trembling, and palpitations and was unable to treat themselves.The customer was provided unspecified treatment by his wife for his symptoms.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key10648986
MDR Text Key210412291
Report Number2954323-2020-06946
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/31/2021
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2021
Initial Date Manufacturer Received 09/30/2020
Initial Date FDA Received10/08/2020
Supplement Dates Manufacturer Received01/22/2021
Supplement Dates FDA Received02/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
Patient Weight101
-
-