GYRUS ACMI, INC M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG Back to Search Results
Model Number M3-30A
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device was received and evaluated.Evaluation determined that the device was found with chipped meniscus causing debris.Fractured lens were observed causing cloudy image.Minor debris underneath the cover glass on eye cup was observed.The identified parts needs to be replaced.The device was placed for repair.
 
Event Description
It was reported that during preparation for use, the device was found with cracked distal tip.There was no patient involvement.No user harm, or injury reported.
 
Manufacturer Narrative
This supplemental report is being submitted to provide review of the device history records (dhr).Please see updated sections: g4, g7, h2, h3,h4, h6 and h10.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The physical damage found to meniscus of the device is a known failure mode, and was most likely due to user mishandling.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states : study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG
Type of Device
M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
MDR Report Key10649140
MDR Text Key210415399
Report Number1519132-2020-00088
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
PMA/PMN Number
K890328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3-30A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-