This supplemental report is being submitted to provide review of the device history records (dhr).Please see updated sections: g4, g7, h2, h3,h4, h6 and h10.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The physical damage found to meniscus of the device is a known failure mode, and was most likely due to user mishandling.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states : study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects.Olympus will continue to monitor complaints for this device.
|