Model Number N/A |
Device Problems
Power Problem (3010); Intermittent Loss of Power (4016)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
It was reported that during surgery the surgeon stated dermatome need more force to take graft.The blade was changed creating a delay in procedure of 31-60 minutes.However the still the same problem existed.Another zimmer dermatome was available to complete the procedure.No impact injury to patient reported.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined that the motor had low rpm.The motor was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|
Event Description
|
No additional event information.
|
|