Model Number 45007 |
Device Problem
Suction Problem (2170)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: date of event was unknown, used the first day in the month it was reported.
|
|
Event Description
|
It was reported that a loss of aspiration occurred.A 2.4mm jetstream xc catheter was selected for a procedure.During the procedure, the catheter lost aspiration.The procedure was completed with another of the same device.There were no patient complications reported.
|
|
Manufacturer Narrative
|
B3: date of event was unknown, used the first day in the month it was reported.Device analysis by mfr: the device was visually examined for any shaft damage and the functional testing of the device was completed.The device showed no damage.Functional analysis was done by completing the setup procedure per the instructions for use.Aspiration testing of the device was done per the test procedure.The device is tested by using a 100ml beaker of water.Test results showed that this device did not perform as designed per the test procedure specification sheet withdrawing 3ml of fluid in the 1 minute time frame.Inspection of the remainder of the device revealed no damage or irregularities.
|
|
Event Description
|
It was reported that a loss of aspiration occurred.A 2.4mm jetstream xc catheter was selected for a procedure.During the procedure, the catheter lost aspiration.The procedure was completed with another of the same device.There were no patient complications reported.
|
|
Search Alerts/Recalls
|