• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETERS Back to Search Results
Catalog Number CRE14S
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2020
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s), and provides general instructions for use, as well as warnings, precautions ,and potential complications associated with the device.Upon receipt of new, or additional information, a follow-up report will be submitted as applicable.The catalog number identified has not been cleared in the u.S., but is similar to the crosser cto recanalization catheters products that are cleared in the us.The pro code and 510 k number for the crosser cto recanalization catheters products are identified.(expiration date: 04/2022).
 
Event Description
It was reported that during a recanalization procedure, the tip of the catheter allegedly detached.There was no patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the crosser cto recanalization catheters products that are cleared in the us.The pro code and 510 k number for the crosser cto recanalization catheters products are identified in d2 and g5.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the alleged deficiency could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiration date: 04/2022), g3 h11: e1 h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device not returned.
 
Event Description
It was reported that during a recanalization procedure, the tip of the catheter was allegedly detached.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETERS
Type of Device
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key10662459
MDR Text Key210878294
Report Number2020394-2020-20021
Device Sequence Number1
Product Code PDU
Combination Product (y/n)N
PMA/PMN Number
K112308
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCRE14S
Device Lot NumberGFEQ3392
Was Device Available for Evaluation? No
Date Manufacturer Received03/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
-
-