Model Number 595000-001 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
|
|
Event Description
|
The customer, a syncardia authorized distributor, reported that the freedom driver connection 1 receptacle was pushed into the freedom driver while supporting a patient.There was no reported adverse patient impact.The customer also reported that the patient was subsequently switched to a backup freedom driver.
|
|
Manufacturer Narrative
|
The customer-reported broken connection to the power adaptor was confirmed via visual inspection where it was observed that the connection 1 receptacle cable was broken and displaced in the driver.This is the connection point between the power adaptor and the driver.The cable was no longer attached to the driver housing and was lodged inside the housing.It cannot be conclusively confirmed how the connection 1 receptacle cable was damaged, but it is most likely the result of either the application of excessive force when connecting the power adaptor to the driver, or rough handling/an impact shock to the driver.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.Ce 5368 follow-up report 1.
|
|