• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE Back to Search Results
Model Number 500101-001
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Insufficient Information (4580)
Event Date 10/01/2020
Event Type  Death  
Manufacturer Narrative
The tah-t was explanted and will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia authorized distributor, reported that the patient passed away on (b)(6) 2020.The customer did not report a cause of death.The customer also reported that the tah-t was explanted and an autopsy was performed.No additional information has been provided at this time.
 
Manufacturer Narrative
The syncardia 70cc temporary total artificial heart (tah-t) is an implantable pulsatile biventricular replacement device that replaces a patient's native ventricles and valves and pumps blood to both the pulmonary and systemic circulation systems.The syncardia 70cc tah-t is indicated for use as a bridge to transplantation in cardiac transplant-eligible candidates at risk of imminent death from biventricular failure.The syncardia tah-t system is intended for use inside and outside the hospital.The explant analysis of the tah-t evaluated the function of the diaphragms, the function and condition of the valves, as well as the overall condition of the heart.Visual inspection of the tah-t found no anomalies or defects.The valve inspection evaluates natural movement to determine if any defects may have been present, preventing them from moving freely as intended.No abnormalities or defects were found with the valves.The diaphragm was visually inspected for any anomalies or defects, as well as tested for strength of stroke volume.Both the left and right ventricle diaphragms were free of defects, and both performed as intended during stroke volume testing.Scoring of the physical condition, functionality, and performance of the tah-t determined that the tah-t was functioning as intended while implanted, and there is no evidence of a device malfunction.Based on the information provided by te customer, and evaluation completed, there is no indication of a device malfunction or performance issue of the device that would impact the reported event.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T)
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key10663381
MDR Text Key211282733
Report Number3003761017-2020-00212
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003008
UDI-Public(01)00858000003008
Combination Product (y/n)N
Reporter Country CodeLO
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Model Number500101-001
Device Catalogue Number500101
Device Lot Number113821
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 10/05/2020
Initial Date FDA Received10/12/2020
Supplement Dates Manufacturer Received10/05/2020
Supplement Dates FDA Received03/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age59 YR
Patient SexFemale
-
-