BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number 4542 |
Device Problems
Failure to Capture (1081); High impedance (1291); Material Integrity Problem (2978)
|
Patient Problem
No Code Available (3191)
|
Event Date 09/30/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
This product has not been returned; therefore, a technical analysis cannot be conducted.Without a returned product it is not possible to definitively confirm how this product may have contributed to the complaint incident.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that during an elective device replacement procedure, this left ventricular lead was intended to be replaced as it had been non functional for approximately two years, due to high out of range pacing impedance measurements and no capture.A procedural x-ray image confirmed a lead conductor fracture near the subclavian area.Another lead was not implanted due to vein occlusion.No other resulting adverse patient effects were observed.
|
|