• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problem Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/03/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under cmp-(b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that during surgery the device was not grafting, it was not cutting properly.There was no harm or delay reported.No further information provided.No adverse event was reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the device was out of calibration at the zero setting so the spring seal and bearings were replaced and device calibrated and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
 
Event Description
No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key10671124
MDR Text Key211461448
Report Number0001526350-2020-00860
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 01/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number63962143
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/07/2020
Initial Date FDA Received10/13/2020
Supplement Dates Manufacturer Received01/06/2021
Supplement Dates FDA Received01/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-