Model Number N/A |
Device Problem
Failure to Cut (2587)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under cmp-(b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that during surgery the device was not grafting, it was not cutting properly.There was no harm or delay reported.No further information provided.No adverse event was reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the device was out of calibration at the zero setting so the spring seal and bearings were replaced and device calibrated and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
|
|
Event Description
|
No additional event information available.
|
|