Model Number N/A |
Device Problems
Failure to Cut (2587); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under cmp-(b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.(b)(6).
|
|
Event Description
|
It was reported that the dermatome did not work properly during surgery.The sound of motor was odd and changing, and then it was noted that the blade does not operate and the graft could not be cut.Other (unidentified) dermatome was used to complete the surgery.There was no procedure delay nor harm/injury to any person involved as a result of this malfunction.
|
|
Event Description
|
It was reported that the dermatome did not work properly during surgery.The sound of motor was odd and changing, and then it was noted that the blade does not operate and the graft could not be cut.Other (unidentified) dermatome was used to complete the surgery.There was no procedure delay nor harm/injury to any person involved as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined that the unit's motor was not functioning.The motor was replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|