H3: device eval by manufacturer: the sentinel cerebral protection system was returned in a generic plastic bag.Overall visual inspection did not identify failures or evidence that could be lost due to the decontamination process.Visual inspection revealed that the sentinel cerebral protection system was returned with the proximal filter sheathed, the articulating distal sheath(ads) relaxed, the distal filter slider (#3) in deployment position, and the distal filter assembly detached from the inner member, which was not returned.The distal section of the inner member was inspected under ultraviolet(uv) light and confirmed presence of adhesive where the distal filter assembly should be.Under microscopic examination, the distal filter assembly detachment was confirmed.The distal section of the inner member was microscopically inspected and no issues were detected.A picture was returned with the sentinel cerebral protection system showing the distal filter of the sentinel cerebral protection system detached from the inner member; additionally, the strut tube is broken and the distal filter is inverted.However, this will not be considered as failures since as per event information, the team intentionally inverted the distal filter when flushing, post-procedure to observe the debris.
|