The field service engineer (fse) followed up with the customer via telephone, and advised customer to power off and on the aia-2000 which resolved reported error.The aia-2000 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review for similar complaints were performed for the serial number (b)(4) from 07jan2019 through aware date 07feb2020.There were no similar complaints identified during the searched period.The aia-2000 operator's manual under section 04 - error messages states the following: [4241] hybrid arm y-axis home detection failure.Cause: the home sensor failed to activate after the hybrid arm y-axis moved toward the home position.If retry fails, the measurement result will be flagged (mf flag).Solution: contact tosoh service center or local representatives.The most probable cause of the reported event is unknown.Submission of this report does not constitute an admission that the manufacturer's product caused, or contributed to the event.
|
Customer reported that they are getting a constant error 4241 hybrid arm y axis home detect.The technical support specialist (tss) had the customer perform an all set home command and the same error persists.The customer switched to their second aia-2000 analyzer to continue their samples.A field service engineer (fse) was dispatched to address the reported issue, which caused a delay in reporting alpha-fetoprotein (afp) patient samples.There was no indication of any patient intervention, or adverse health consequences due to the delay in reporting of patient results.
|