• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION AIA-2000

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION AIA-2000 Back to Search Results
Model Number AIA-2000
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2020
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) followed up with the customer via telephone, and advised customer to power off and on the aia-2000 which resolved reported error.The aia-2000 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review for similar complaints were performed for the serial number (b)(4) from 07jan2019 through aware date 07feb2020.There were no similar complaints identified during the searched period.The aia-2000 operator's manual under section 04 - error messages states the following: [4241] hybrid arm y-axis home detection failure.Cause: the home sensor failed to activate after the hybrid arm y-axis moved toward the home position.If retry fails, the measurement result will be flagged (mf flag).Solution: contact tosoh service center or local representatives.The most probable cause of the reported event is unknown.Submission of this report does not constitute an admission that the manufacturer's product caused, or contributed to the event.
 
Event Description
Customer reported that they are getting a constant error 4241 hybrid arm y axis home detect.The technical support specialist (tss) had the customer perform an all set home command and the same error persists.The customer switched to their second aia-2000 analyzer to continue their samples.A field service engineer (fse) was dispatched to address the reported issue, which caused a delay in reporting alpha-fetoprotein (afp) patient samples.There was no indication of any patient intervention, or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
AIA-2000
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
bernadette oconnell
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586-23
JA   1058623
MDR Report Key10671806
MDR Text Key213501635
Report Number8031673-2020-00298
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000
Device Catalogue Number022101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2020
Initial Date FDA Received10/13/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-