• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA (SH) TRIFECTA GT VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA (SH) TRIFECTA GT VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TFGT-23A
Device Problems Backflow (1064); Material Split, Cut or Torn (4008)
Patient Problems Aortic Regurgitation (1716); Insufficiency, Valvular (1926)
Event Date 09/10/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
On (b)(6) 2018, a 23mm trifecta gt valve was implanted as a bentall in a graft.On (b)(6) 2020, a valve in valve procedure was performed due to a leaflet tear that caused high aortic insufficiency.No adverse health consequences were reported.
 
Manufacturer Narrative
An event of a leaflet tear and high aortic insufficiency was reported.A more comprehensive assessment could not be performed as a valve-in-valve procedure was performed and the device was not accessible for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIFECTA GT VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA (SH)
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 01897 -405
CS  01897-4050
MDR Report Key10671993
MDR Text Key211257295
Report Number3008452825-2020-00554
Device Sequence Number1
Product Code LWR
UDI-Device Identifier05415067018229
UDI-Public05415067018229
Combination Product (y/n)N
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/27/2022
Device Model NumberTFGT-23A
Device Catalogue NumberTFGT-23A
Device Lot Number6368413
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/16/2020
Initial Date FDA Received10/13/2020
Supplement Dates Manufacturer Received10/15/2020
Supplement Dates FDA Received10/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-