• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: QUEST MEDICAL, INC. MPS 2 CONSOLE; CPBP HEAT EXHANGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUEST MEDICAL, INC. MPS 2 CONSOLE; CPBP HEAT EXHANGER Back to Search Results
Model Number 5201260
Device Problems Electrical /Electronic Property Problem (1198); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2020
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned to quest medical for evaluation.A follow-up medwatch will be submitted if additional information becomes available.
 
Event Description
A report was received regarding an alleged issue encountered with the console.The report states that the console displayed ec179 during bypass.The console was swapped out and the case was resumed.There were no patient complications.
 
Manufacturer Narrative
The console was evaluated and the resistor in the motor driver board was identified as the cause of the reported issue.The motor driver board was replaced and the console passed all operational verification tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MPS 2 CONSOLE
Type of Device
CPBP HEAT EXHANGER
Manufacturer (Section D)
QUEST MEDICAL, INC.
one allentown parkway
allen TX 75002
MDR Report Key10674376
MDR Text Key213199104
Report Number1649914-2020-00036
Device Sequence Number1
Product Code DTR
Combination Product (y/n)N
PMA/PMN Number
K173716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number5201260
Date Manufacturer Received09/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-