• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 52MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 52MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120152
Device Problem Compatibility Problem (2960)
Patient Problem Injury (2348)
Event Date 03/19/2019
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed on (b)(6) 2019 on the patient right hip due to pain, limited mobility, metallosis, tissue damage and elevated metal ion levels.The patient outcome is unknown.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr head and bhr cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr head.Similar complaints have been identified for the bhr cup.However, as bhr systems of this size are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.The clinical information provided, the report of elevated metal ion levels and ¿mild staining¿ may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.It is also unknown if this patient¿s documented ¿paraproteinemia¿ contributed to the clinical symptoms suffered by this patient.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 52MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key10675643
MDR Text Key211221354
Report Number3005975929-2020-00385
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502575
UDI-Public3596010502575
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number74120152
Device Lot Number077994
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/18/2020
Initial Date FDA Received10/13/2020
Supplement Dates Manufacturer Received02/04/2021
Supplement Dates FDA Received02/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-