SMITH & NEPHEW, INC. GII OVAL PATELLA 5 PT. SZ GDE; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
Model Number 71440842 |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/16/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the genesis ii oval patella 5 pt.Size gde was returned with one of the arms broken.No information about surgical delays, s+n backup devices or when this incident happened was initially provided.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Manufacturer Narrative
|
H10: internal complaint reference: (b)(4).H2: additional information ¿section¿ h3, h6: ¿the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The 38mm arm of the device was found to be fractured from the device, rendering the device inoperative.The fractured component was returned as well.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.¿.
|
|
Search Alerts/Recalls
|
|
|