ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/29/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of battery sn (b)(4) has been completed.The reported problem (will not power on monitor) has been confirmed.As received, the battery was unable to power on a monitor or charge.Upon evaluation, the battery pack tri-cells were depleted.The battery did not last 24 hours before being discharged.The root cause of the depleted cells was unable to be positively identified.No adverse events occurred as a result of the defective battery.
 
Event Description
(b)(6) 2020: resubmitting this mdr as part of an internal audit where the electronic 3500a pdf form was located, and acknowledgements 1, 2, and 3 could not be located.A us distributor contacted zoll to report that a battery was unable to power on a monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key10676997
MDR Text Key211505194
Report Number3008642652-2018-00699
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005036
UDI-Public00855778005036
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 01/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0989
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2017
Initial Date Manufacturer Received 01/15/2018
Initial Date FDA Received10/14/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-