Device evaluation of monitor sn (b)(4) has been competed.The reported problem (check te pad messages, discharge profile faults) was confirmed.As received, the monitor failed incoming testing, and produced service code 105 and service code 204.Upon evaluation, the q12 and q16 transistors were physically damaged.In addition, the inter-pca connectors j1000, j1002, and j1003 were defective.The cause for the failure is the defective components.In addition, there was extensive physical damage found to the monitor.The root cause of the defective components cannot be positively identified.No adverse events occurred as a result of the defective monitor.
|