Model Number WCD 4000
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) has been competed.The reported problem (check te pad messages, discharge profile faults) was confirmed.As received, the monitor failed incoming testing, and produced service code 105 and service code 204.Upon evaluation, the q12 and q16 transistors were physically damaged.In addition, the inter-pca connectors j1000, j1002, and j1003 were defective.The cause for the failure is the defective components.In addition, there was extensive physical damage found to the monitor.The root cause of the defective components cannot be positively identified.No adverse events occurred as a result of the defective monitor.
 
Event Description
A u.S.Distributor returned monitor sn (b)(4), and reported that the device caused, "check therapy pad" messages, and discharge profile faults.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key10677007
MDR Text Key211349154
Report Number3008642652-2018-00726
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2020
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0988
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Date Manufacturer Received01/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-