MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT3 III; RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT3 III; RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE Back to Search Results
Model Number FT3 G3
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/17/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).The investigation is ongoing.
 
Event Description
The initial reporter received questionable elecsys ft3 iii results for 1 patient sample on a cobas e 801 module, serial number (b)(4).This medwatch covers the alleged results for the ft3 assay.Please refer to medwatch mdr-47440 for the alleged tsh assay results and mdr-47441 for the alleged ft4 assay results.(b)(4).The results were reported to the patient's physician, who requested the sample test be repeated.
 
Manufacturer Narrative
The patient's sample was returned for investigation.Based on the results generated, the investigation determined an interfering factor was present in the sample.The different results are consistent with a physical reaction to the ft3 assay and the effect of the interfering factor.Product labeling states: "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings." the investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT3 III
Type of Device
RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10678532
MDR Text Key211499140
Report Number1823260-2020-02545
Device Sequence Number1
Product Code CDP
Combination Product (y/n)N
PMA/PMN Number
K963127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT3 G3
Device Catalogue Number07027362190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/18/2020
Initial Date FDA Received10/14/2020
Supplement Dates Manufacturer Received09/18/2020
Supplement Dates FDA Received12/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-