Model Number WCD 4000
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/22/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of battery pack sn (b)(4) has been completed.The reported problem (unable to power up a monitor) was confirmed.As received, the battery pack was unable to power a test monitor and charge.The cause for the failure was isolated to a loose bus bar inside of the battery.The root cause of the loose busbar cannot be positively identified.No adverse events occurred due to the defective battery.
 
Event Description
A us distributor reported that a battery would not power on a monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
anna beatty
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key10678585
MDR Text Key213256054
Report Number3008642652-2020-08938
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005036
UDI-Public00855778005036
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0989
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Initial Date Manufacturer Received 10/13/2020
Initial Date FDA Received10/14/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-