Model Number AS480
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2020
Event Type  malfunction  
Manufacturer Narrative
Root cause: potential alteration of staining in this case was due to improper operation of the compressor.The problem was solved by field service engineer with replacement of the part.Following the replacement, the instrument was fully operational within specifications, without errors and available for the user.Failure mode description: failure of the compressor, may result in false negative, background staining or uneven (weak, inconsistent) staining.This failure could be due to broken parts or malfunction causing air blow to not prepare all slides for dispense.The compressor failure mode described above has the potential to alter staining.
 
Event Description
Based on complaint report or investigated failure mode, there was a staining alteration.Customer complaint record reported the event as follows: weak/inconsistent staining.No direct or indirect patient harm or user harm have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSTAINER LINK 48
Type of Device
AUTOSTAINER LINK 48
Manufacturer (Section D)
DAKO NORTH AMERICA, INC.
6392 via real
carpinteria CA 93013
Manufacturer (Section G)
DAKO NORTH AMERICA, INC.
6392 via real
carpinteria CA 93013
Manufacturer Contact
benjamin gaither
6392 via real
carpinteria, CA 93013
5123328207
MDR Report Key10679322
MDR Text Key211459111
Report Number2022180-2020-00356
Device Sequence Number1
Product Code KPA
UDI-Device Identifier05700572035497
UDI-Public05700572035497
Combination Product (y/n)Y
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberAS480
Device Catalogue NumberAS48030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/14/2020
Initial Date FDA Received10/14/2020
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-