Catalog Number JC7751 |
Device Problem
Failure to Prime (1492)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the filter chamber of a y-type blood/soln set with standard blood filter would not prime; this was further described as ¿when they attempted to prime the tubing, the filter chamber would not fill with priming fluid¿.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added to d10, h3 and h6.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional test including pressure test and clear passage test was performed and no leaks or blockages were noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information was added to h4 and h6.H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|