• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number JC7751
Device Problem Failure to Prime (1492)
Patient Problem No Patient Involvement (2645)
Event Date 08/14/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the filter chamber of a y-type blood/soln set with standard blood filter would not prime; this was further described as ¿when they attempted to prime the tubing, the filter chamber would not fill with priming fluid¿.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d10, h3 and h6.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional test including pressure test and clear passage test was performed and no leaks or blockages were noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to h4 and h6.H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10680273
MDR Text Key211438949
Report Number1416980-2020-06417
Device Sequence Number1
Product Code BRZ
UDI-Device Identifier00085412152738
UDI-Public(01)00085412152738
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJC7751
Device Lot NumberDR20E17025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Date Manufacturer Received12/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-