Model Number G140 |
Device Problems
High impedance (1291); Material Integrity Problem (2978)
|
Patient Problems
No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated.
|
|
Event Description
|
It was reported during a revision procedure that this cardiac resynchronization therapy defibrillator (crt-d) device exhibited high out of range pacing impedance measurements of 3,000 ohms.A different device was implanted.Testing of the right ventricular (rv) lead with the pacing system analyzer (psa) confirmed this was not a lead issue, and more than likely a header issue.No adverse patient effects were reported.The device is expected to be returned for analysis.
|
|
Event Description
|
It was reported during a revision procedure that this cardiac resynchronization therapy defibrillator (crt-d) device exhibited high out of range pacing impedance measurements of 3,000 ohms.A different device was implanted.Testing of the right ventricular (rv) lead with the pacing system analyzer (psa) confirming this was not a lead issue, and more than likely a header issue.No adverse patient effects were reported.The device is expected to be returned for analysis.
|
|
Manufacturer Narrative
|
The returned device was inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.Pacing and sensing functions were tested and the device was verified to operate as expected.Analysis did not identify any device characteristics that would have caused or contributed to the report clinical observations.
|
|
Search Alerts/Recalls
|