SMITH & NEPHEW, INC. GII PRIM P/S COLLET W/THMBSCRW; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71440467 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/16/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during the procedure, the gold wheel broke when unscrewing it from the block.The procedure was completed without delay and backup from smith&nephew was available to finish the surgery.No patient injury/harm or other complications at the moment.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The device, used in treatment, was returned for evaluation.A visual inspection of the returned collet confirms the thumbnut is broken.All the pieces of the device were returned for evaluation.The device shows extreme signs of wear/usage.The device was manufactured in 2001.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|