• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PRIM P/S COLLET W/THMBSCRW; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PRIM P/S COLLET W/THMBSCRW; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71440467
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/16/2020
Event Type  malfunction  
Event Description
It was reported that during the procedure, the gold wheel broke when unscrewing it from the block.The procedure was completed without delay and backup from smith&nephew was available to finish the surgery.No patient injury/harm or other complications at the moment.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned collet confirms the thumbnut is broken.All the pieces of the device were returned for evaluation.The device shows extreme signs of wear/usage.The device was manufactured in 2001.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PRIM P/S COLLET W/THMBSCRW
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10681834
MDR Text Key211435562
Report Number1020279-2020-05407
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010443076
UDI-Public03596010443076
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/27/2011
Device Model Number71440467
Device Catalogue Number71440467
Device Lot Number01LM03823
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-