It was reported that the patient presented in clinic for follow-up.Upon review, the patient's pulse generator exhibited no left ventricular capture and a rise in impedance.The device also exhibited increased atrial capture thresholds, increased impedance and lead noise.The patient was screened and the lead pins did not appear to be properly inserted in the header.On (b)(6) 2020 the patient's presented for a revision procedure.During the procedure, the physician noted that they were unable to insert the pins any further than they were originally.The patient's device was removed and replaced.Both the right atrial and left ventricular leads tested normally and were reused with the new device without issue.The patient is recovering.
|