Model Number 1456Q/86
Device Problems Difficult to Insert (1316); Separation Failure (2547)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/28/2020
Event Type  malfunction  
Manufacturer Narrative
This product is registered as a combination product.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported the asymptomatic patient presented for an implant procedure.During surgery, the lead was very difficult to get through the sheath.During the slitting, the cutter tool came out of the sheath and the lead was kinked.The guidewire also got stuck and the lead was replaced.The procedure was completed successfully.The patient was stable.
 
Manufacturer Narrative
A complete lead was returned in one piece with a guide wire inserted and was measured to be 86.0 cm.Analysis concluded the reported events of difficult to insert and guidewire separation failure were confirmed.The damage was consistent with procedure damage.The lead was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10687590
MDR Text Key211649452
Report Number2017865-2020-15598
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000089789
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age78 YR
Patient Weight100
-
-