Device Problem Entrapment of Device (1212)
Patient Problems Cardiac Arrest (1762); Foreign Body In Patient (2687)
Event Date 09/22/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The instructions for use for the peripheral orbital atherectomy system states: "the diamondback 360® peripheral orbital atherectomy system is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy." use of the system is contraindicated if the target lesion is within a bypass graft or stent.The device history record for the reported oad was unable to be reviewed, as the lot number was not provided.(b)(4).
 
Event Description
A 1.25 solid crown peripheral orbital atherectomy device was used to treat a chronic total occlusion in a previously placed stent in the left anterior descending artery.The crown of the oad became stuck in the stent and could not be removed using balloon angioplasty, snare, or microcatheter.The shaft of the device was cut, and the crown remained in the patient.The patient coded during the procedure and was placed on extracorporeal membrane oxygenation (ecmo).As of (b)(6) 2020, the patient had been removed from ecmo and was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Type of Device
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
st. paul, mn MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
st. paul, mn MN 55112
Manufacturer Contact
laramie otto
1225 old highway 8 nw
st. paul, mn, MN 55112
6512591600
MDR Report Key10691275
MDR Text Key211720274
Report Number3004742232-2020-00322
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/22/2020
Initial Date FDA Received10/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age39 YR
-
-