SMITH & NEPHEW, INC. GII LONG I/M ROD 16 INCH; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71440004 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/21/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the proximal end of the rod broke off when removing.It occurred during use, with instruments outside the patient.No pieces fell inside the patient and all pieces were recovered.No patient injury or surgical delay was reported.The procedure was completed with the same device.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The end of the device is fractured off, the fractured piece was returned, rendering the device inoperable.The device was manufactured in 2006 and shows signs of extensive use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.H4: no valid manufacture date.
|
|
Search Alerts/Recalls
|
|
|