• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII MIS DCF ALIGN GDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII MIS DCF ALIGN GDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71441144
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Patient Involvement (2645)
Event Date 09/28/2020
Event Type  malfunction  
Event Description
It was reported that during inspection it was found that the cutting block doesn't stay locked in place.No case involved.
 
Manufacturer Narrative
The associated device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MIS DCF ALIGN GDE
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10697705
MDR Text Key211902903
Report Number1020279-2020-05504
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010497161
UDI-Public03596010497161
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441144
Device Catalogue Number71441144
Device Lot Number17EM15091
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CN: 71441147, LN: 15GM04009; CN: 71441147, LN: 15GM04009
-
-