• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD COMPOSITE SERIES SKULL CLAMP; ¿COMPOSITE SERIES¿

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD COMPOSITE SERIES SKULL CLAMP; ¿COMPOSITE SERIES¿ Back to Search Results
Catalog Number A3059
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
It was reported that the locking mechanism of the a3059 mayfield composite series skull clamp slips or was broken.There was no known patient injury or surgery delay.
 
Manufacturer Narrative
Unique device identifier (udi): (b)(4).The mayfield skull clamp was returned for evaluation: device history record (dhr) - no abnormalities related to the reported failure.Failure analysis - no issues observed.With respect to the returned unit, it has passed all specific functional testing requirements.Evaluation found when the unit is properly positioned and put under pressure, the unit would not have slipped.The reported complaint was not confirmed.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10697919
MDR Text Key212142960
Report Number3004608878-2020-00618
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2020
Date Manufacturer Received10/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-