MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 UNKNOWN HIP FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 UNKNOWN HIP FEMORAL HEAD Back to Search Results
Catalog Number UNK HIP FEMORAL HEAD
Device Problems Difficult to Insert (1316); Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Test Result (2695); No Code Available (3191); Unspecified Tissue Injury (4559)
Event Date 10/03/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Initial reporter occupation: lawyer.(b)(4).
 
Event Description
Asr litigation records received.Litigation alleges injury, elevated chromium and cobalt levels, economic loss, pain, suffering, and emotional distress.Doi: (b)(6) 2007; dor: (b)(6) 2019; right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
After review of medical records, the patient was revised due to failed right thr and pain.It was also indicated that there was a surgical delay of 3 hours due to the fact that a well-fixed femoral stem had to be removed that required complex intraoperative decision making.Operative notes reported that it was apparent during reduction of the hip that the femoral head was actually moving on to the trunnion.There was marked metal debris in this region and fair amount of metallic looking fluid was encountered.The trunnion was worn in both lateral and anterior-posterior direction.Metallic debris and but necrotic tissues were removed.Doi: (b)(6) 2007, dor: (b)(6) 2019, right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL HEAD
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key10699345
MDR Text Key212085774
Report Number1818910-2020-22781
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/05/2020
Initial Date FDA Received10/19/2020
Supplement Dates Manufacturer Received01/07/2021
Supplement Dates FDA Received01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
ASR; ASR; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-