Model Number 199721001
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/04/2020
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2020 during a transforaminal lumbar interbody fusion (tlif) procedure, the surgeon was using the expedium verse spine system.During final tightening, the surgeon heard a click that was not the usual torque limiting click sound.They used the correction key locking caps.The set screw was still spinning freely after the cap was attached and fully seated in the drive mechanism.The surgeon was backing the screw out when metal shards came off from the screw tulip.This occurred with six (6) screws.There was a surgical delay of sixty (60) minutes.The procedure was successfully completed.There was no patient consequence.This report is for one (1) 5.5 exp verse unitized set scr.This is report 3 of 5 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11: corrected data: g4: awareness date reported on initial report as september 04, 2019 but should have been october 16, 2020.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h4, h6- the 5.5 exp verse unitized set scr (p/n: 199721001, lot number: xd2132) was received at us cq.Visual inspection of the complaint device showed that the threads of the verse unitized set screw were broken.The broken fragments were not returned.No other issues were identified.This complaint can be confirmed for 5.5 exp verse unitized set scr (p/n: 199721001, lot number: xd2132).No definitive root cause can be established.However, the broken threads could have been a result of unintended forces applied to the device.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the dhr of product code 199721001, lot xd2132, was reviewed and no non-conformance was observed during the manufacturing process.The product was released on march 31, 2020.The dhr was electronically reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10: additional narrative: d10, d11, h3, h6: the device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Concomitant device reported: final tightener (part unknown, lot unknown, quantity unknown); rod (part unknown, lot unknown, quantity unknown).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: h4 device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE UNITIZED SET SCR
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10700967
MDR Text Key229282654
Report Number1526439-2020-02000
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034466156
UDI-Public(01)10705034466156
Combination Product (y/n)N
PMA/PMN Number
K142185
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 09/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number199721001
Device Catalogue Number199721001
Device Lot NumberXD2132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2020
Date Manufacturer Received10/30/2020
Patient Sequence Number1
Treatment
5.5 EXP VERSE UNITIZED SET SCR; 5.5 EXP VERSE UNITIZED SET SCR; 5.5 EXP VERSE UNITIZED SET SCR; UNKNOWN RODS; UNKNOWN TIGHTENER; VERSE CORRECTION KEY; VERSE CORRECTION KEY; VERSE CORRECTION KEY; VERSE CORRECTION KEY; VERSE CORRECTION KEY
-
-