Catalog Number J2C1711
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/24/2020
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a small volume infusor leaked from the bladder or tubing (not further clarified) during patient infusion.There was damage on the surface of the housing and coil cap.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H4: the lot was manufactured from march 24, 2020 - march 25, 2020.H10: the device was received for evaluation with no fluid in the bladder.Visual inspection was performed using the naked eye which observed dents on the housing and the pink coil cap.Additionally, a scrape was observed on the tubing line.Functional testing was performed by filling the device with green colored water.During fill, a leak was observed from the scrape on the tubing line only.There was no leak observed from any other part of the device.The reported condition was verified.The cause of the condition was due to user induced damage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10701318
MDR Text Key212108201
Report Number1416980-2020-06489
Device Sequence Number1
Product Code MEB
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJ2C1711
Device Lot Number20C010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2020
Initial Date Manufacturer Received 09/24/2020
Initial Date FDA Received10/19/2020
Supplement Dates Manufacturer Received10/26/2020
Supplement Dates FDA Received11/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-